Device Setup
This page provides details of how devices are setup ready to be sent to participants of the SAFER Wearables Study.
Table of contents
- Cleaning
- Bittium Faros 180
- PulseOn OHR Tracker
- PulseOn Arrythmia Monitor
- Withings Move-ECG
- Mobile Phone
Cleaning
All devices should be cleaned using wipes.
Bittium Faros 180
- Connect the device to a Windows PC via USB.
- Delete the ‘DATA’ folder from the device.
- Open Bittium Faros Manager
- Keep the device attached until it is fully charged (this could take 1-2 hours), as indicated by 100% ‘Battery Charge’ in the Bittium Faros Manager software.
- In Bittium Faros Manager, make sure that the device is setup correctly:
- Check that the device name is ‘SN_#######’, where ####### is the device serial number.
- Check that the date and time are coorect and that the device time is set to synchronize with PC time.
- ECG/EKG set to ‘On’
- Accelerometer set to ‘On’
- RR-interval set to ‘Off’
- Number of ECG channels set to ‘1’ (there isn’t another option)
- ECG Sampling Frequency set to ‘500 Hz’
- Accelerometer Sampling Frequency set to ‘25’.
- Accelerometer Dynamic Range set to ‘4g’.
- Autostart on Skin Contact set to ‘On’ [TBC].
- Press SAVE to store these settings.
- In addition, make the following manual adjustments to the
.\SYS\FAROS.CFG
file (which is a text file):- bat_alarm_beep set to ‘0’
- Save the file after making these adjustments
- Check that the device doesn’t have any data stored on it (by using Windows File Explorer to inspect the
.\DATA
folder on the device. Delete any files in this folder. This should leave approximately 93% ‘Free Memory’, as stated in Bittium Faros Manager. - Use Bittium Faros Manager to eject the device.
- Unplug.
- Check that the device is turned off (there are no lights showing, even intermittently, on the device).
- Insert the electrode set into the USB port of the device.
PulseOn OHR Tracker
- Sync the device with a mobile phone using the ‘PulseOn Demo’ application
- Charge the device via USB until it is fully charged (as indicated by the battery indicator reaching the top of the outer rectangle in the battery logo).
- Select ‘disconnect’ in the ‘PulseOn Demo’ application.
- Unplug, put back in the wristband.
- Close the ‘PulseOn Demo’ application.
- Sync the device once more
- Set the device mode (used ‘Sampled mode’ ‘HR’ until 09-July-2024, now using ‘Continuous mode’, ‘HR+RAW’ from 10-July-2024 onwards) in the ‘PulseOn Demo’ application
- Select ‘disconnect’ in the ‘PulseOn Demo’ application.
- Turn the device off (by holding the button for 10 secs), and it is ready to send.
PulseOn Arrythmia Monitor
- Charge the device via USB until it is fully charged (as indicated by the battery indicator being green).
- Follow the instructions in the README.md file provided by PulseOn, using:
- https://prod.pulseon-ecg.uk/
- device serial number “SN#####” as the first name, “SETUPYYYYMMDD” as the last name, and “SN#####SETUPYYYYMMDD” as the identifier
- The following settings for the measurement session: time zone - “00:00 Europe/London”; (i) session type - “Screening (without gateway)”; (ii) session length - “1 weeks”; (iii) quiet time - “22:00-08:00”; (iv) timed measurements - “09:00”, “12:00”, “16:00”, “20:00”;
Withings Move-ECG
- Open HealthMate on a mobile phone.
- …
Mobile Phone
Initial setup
- Initial turning on:
- select ‘English (UK) language;
- connect to WiFi;
- allow to check for updates;
- don’t copy apps and data;
- don’t sign in to Google;
- in Google Services settings turn off ‘use location’, ‘allow scanning’, ‘send usage and diagnostic data’ options, and agree to terms of service;
- select Google as search engine;
- skip setting screen lock;
- under ‘Review additional apps’ press ok;
- under ‘About your privacy’, select ‘Privacy settings’, and turn off ‘Help improve Motorola products’ and ‘Enhanced device support’, and press ‘Accept and continue’;
- don’t enter email address;
- don’t grant location permission.
- Settings:
- Sound: turn all four volumes down to minimum; under advanced settings turn off all options except ‘Charging sounds and vibration’, and ‘Touch vibration’.
- Display: Set ‘Brightness level’ to about 75%; under advanced settings set ‘Screen timeout’ to 5 minutes, ‘Font size’ to largest, ‘Display size’ to largest, ‘Screen saver’ to never start
-
Home screen: Remove all apps on screen, including those on the bottom bar;
- Installing apps:
- Play Store: Login using Google Account, turning off ‘Back up to Google Drive’ when the option appears. Install ‘Withings Health Mate’, and ‘MacroDroid’.
- Withings Health Mate: Login; install Move ECG device (including allowing access to location, and setting daily step goal to 15,000, and selecting the ‘Right’ wrist); turn off persistent notifications under settings.
- PulseOn Demo: Install the PulseOn Demo app using the ‘Application-pulseonDemo-1.2.0-debug.apk’ file (allowing required permissions). NB: the macro settings may need updating when using large font.
- MacroDroid: Share the following macros from an already setup phone via Bluetooth: (i) Cancel macros; (ii) Close all apps; (iii) Launch healthmate; (iv) Launch pulseon software. Then use the ‘Export/Import’ function in MacroDroid to import each one in turn from the ‘Bluetooth’ folder, giving the necessary permissions. Finally, add the red, green and yellow MacroDroid widget buttons on the home screen.
- Finishing setup:
- Sign out of Google account
- Deny location access to all apps except for ‘Health Mate’ and ‘PulseOn Demo’, and for these, set to ‘Allowed only while in use’.
- Add HealthMate and PulseOn apps to bar at bottom of screen.
- Add MacroDroid app at top of screen
- Add label to back of phone indicating power button
- Add protective cover